RVP Pro

The RVP Pro performs vapor pressure analysis of crude oil, gasoline, jet fuel and solvents. Its industry proven measuring cell employs sample valving that offers unmatched rinsing of the entire sample circuit, nearly eliminating sample carryover. RVP Pro significantly outperforms the published precision of all available standard test methods.

Description

PSSC introduces the RVP Pro, a vapor pressure analyzer designed to meet the rigorous demands of high-volume laboratories and field operations. RVP Pro offers professional accuracy and precision, simplicity, and durability while offering the most user-friendly experience and comfort. The RVP Pro has a pressure range of 0-1000 kPa, industrial grade hardware and electronics, and unmatched portability. The RVP Pro has applications covered from oil field to fuel pump. ISO 17025 accredited calibration is standard with all Pro deliveries!

Most comfortable and user-friendly vapor pressure analyzer. One button operation with simple user profile and method creation. Engineered and built for years of reliable vapor pressure measurements, using the most robust components, Industry proven measuring cell and valving technology, Transport lock for securing the measuring cell shaker system during shipping and field transport. Self-diagnostics and data logging. Linux OS for fast and easy Software upgrades – for full life of analyzer

The peripheral interface supports LIMS data transfer, ticket printers, full page printers, USB mouse, USB keyboard, USB storage, and barcode scanners, No external PC required, ever.

Designed, built, and serviced in the US

Standard Test Methods

ASTM D5188, ASTM D5191, ASTM D6377, ASTM D6378

Excellent correlation to ASTM D323 (RVPE) and ASTM D4953 (DVPE)

4 x user-programmable methods

Configurable for international vapor pressure methods: EN 13016-1, EN 13016-2, IP 394, IP 409, IP 481, JIS K2258-2, SHT 0769, GOST 52340

Applicable to Fuel Specifications: ASTM  D910,  ASTM1655,  ASTM  D4814,  ASTM D6227

Specifications

Temperature Range: 0 to 100° C (32 to 212° F)

Temperature Stability: +/- 0.01° C (0.02 °F)

Temperature Profiles: Celsius, Fahrenheit

Pressure Range: 0 to 1000 kPa (0-145 psi)

Pressure Resolution: 0.1 kPa (0.07 psi)

Pressure Scales: kPa, hPa, psi, atm

Vapor/Liquid Ratio: 0.2/1 to 100/1, user-adjustable for D6377 and freely programmable methods

Measurement Time: 5 minutes for standard measurement time (user programmable in free methods)

Sample Volume: 1 mL sample, 2.5 mL rinsing volume for RVP methods, 0.25 mL vapor liquid ratio

Sample Introduction: Automatic with built in piston by siphon tube or syringe. Serviceable/replaceable inlet filter (spares inc.)

Sample Conditioning: Built-in shaker with user selectable shaking speed (1.0 – 5.0 c/s) for D6377, D5188, and 4x custom (user programmable) methods Programmable Injection temperature for D6377 and custom methods

Results Database: 50,000+ detailed measurement results

Results Format: csv, rtf, xml

Inputs: Capacitive touch screen, USB mouse, USB keyboard, bar code reader

Outputs: LIMS, Network/Ticket/full-page printer, USB drive

Operating System: Linux

Display: Industrial grade, 8.6″ color touchscreen

Comm. Interfaces: 3x USB type A, 1x USB type B, RS232, Ethernet 10/100 Mbit/s compliant with IEEE802.3-2002 standard

Languages: English, Spanish, German, French

Power Supply: Auto detect 100-240 VAC, 50/60 Hz

Dimensions: H 400 x 300 x 300 mm (15.5 x 11.5 x 11.5 inches)

Weight 11.4 Kg (25 lbs.)

Accessories

RVP Pro is delivered with a custom-designed Pelican case for North American customers that offers maximum protection when transporting to the testing site or shipping the analyzer for service and calibration, ISO/IEC 17025-2017 calibration certificate, USB drive with operating manual, quick start card, 10mL glass syringe, waster container, discharge hose, luer inlet sipper tube, power cord, piston oil, 1 mL plastic syringe

Download brochure